Proper Use; Unpacking; Commisioning; Starting - IKA IKAMAG RCT basic Mode D'emploi

Masquer les pouces Voir aussi pour IKAMAG RCT basic:
Table des Matières

Publicité

Proper use

The IKAMAG RCT basic is a stirring instrument with heating func-
tion. It is used in e.g. laboratories of the chemical industry, in
scools as well as in pharmacies. The instrument is suitable for
tempering substaces which are placed on the heating plate in ves-
sels. By means of a built-in stirring drive and with the help of a
magnetic bar within the vessel, the substances can be stirred at
the same time. The mixing intensity is dependent on the motor
speed and on the size of the magnetic bar.

Unpacking

Please unpack the equipment carefully and check for any dama-
ges. It is important that any damages which may have arisen
during transport are ascertained when unpacking. If applicable a
fact report must be set immediately (post, rail or forwarder).
The delivery scope covers: an IKAMAG RCT basic and operating
instructions.
Commissioning
Check whether the voltage specified on the type plate matches
the mains voltage available. The power socket used must be ear-
thed (protective earth conductor contact). If these conditions are
met, the device is ready to operate after plugging in the mains
plug. If these procedures are not followed, safe operation cannot
be guaranteed and/or the equipment may be damaged.
Observe the ambient conditions (temperature, humidity, etc.)
listed under Technical Data.
10

Starting

After th instrument has been switched on, it is in „standby"-ope-
ration. The functions „heating" and „stirring" are started with the
help of switches on the left and right bottom of the housing
Safe Temp
On - Off
Temp °C
Temp
1
50
1
00
2
00
2
50
50
Schalter
switch
LED
LED
Skala
dial
Drehknopf
rotary
LED
LED
Heizen
Temp
Temp
Temp
Mot
„heating"
temp
temp
knob temp
motor

Function Heating

The heating plate temperature of the instrument is kept at a
constant temperature by the control circuit. The heating plate
temperature is also monitored by a separate adjustable safety cir-
cuit. The temperature sensors required for this (two thermo ele-
ments) are integrated into the heating plate. The instrument dis-
poses of a silumin heating plate with a heating power of 600
watts. The surface of the heating plate is polished in order to faci-
litate cleaning. This surface gets coated rather quickly with an
aluminium oxide protective layer which is chemically very resisi-
tend.
tion „heating" is switched on and/or off by its corresponding
switch. When the function „heating" is switched on, the LED
„temp" shines green. The temperature of the heating plate is set
on the rotary knob „temp" with its corresponding dial (0-300°C).
(see illustration 1)
illustration 1
IKA
®
RCT basic
safety control
On - Off
Mot
Mot
5
6
4
7
3
8
2
9
1
10
dial
Skala
rotary
Drehknopf
Schalter
switch
Mot
Rühren
Mot
motor
knob motor
„stirring"
The func-
RCT B sc 0503

Publicité

Table des Matières
loading

Table des Matières