Define the Event (alarm ) High alarm H for Channel 1, 2 & 3, then define Job1 with Log
alarm. Then, do the rest of the settings according to 4.2 Display, 4.3 Tools, 4.4
Instrument and 4.5 Clock. If the IP address of the recorder is fixed for Ethernet
communication, (not automatically assigned by the server), then the user needs to
define IP as User Defined. Then set up the IP address and Subnet Mask in 4.4
Instrument. The recorder is now ready for real-time monitoring and data acquisition.
If the user needs no communication and wishes to read measured data from the CF
card on a PC, then set up the basic software Observer I in the PC, and add a CF card
reader on USB port of PC. To read the configuration and measured data on CF
card for the first time, it is necessary to press the Save key to save configurations
first , and then press the Dump key to Dump measured data from the recorder to CF
card before inserting it into the CF card reader.
changed, it is necessary to press both keys before inserting the CF card into the
reader . If the configuration remains unchanged, to dump measured data press the
Dump key, do not press Save key again. After that, the measured data can be seen in
trend or Excel format on PC. A 16 MB CF card is supplied free.
5. PC software – Observer I & II
Observer I & Observer II are PC software to be used with the recorder. They are
Windows 98, NT, 2000 and XP compatible to archive, analyze measured data from the
recorder and to do recorder configuration through PC. Their different applications are,
CA650 without communication : Observer I Basic software, supplied free for the
CA650 with communication : Observer II
5.1.

Hardware requirements

It is recommended to use 200 MHz Pentium PC with memory 64 MB RAM.
performance will be compromised if a slower PC is used, or with less memory.
5.2.

Set up

To set up the software on PC, insert the CD, double click the Setup program and follow
the step by step instructions.
5.3.
PC Software manual – HELP
After completing the Set up procedure, the user may get into HELP on the PC
software and use it as the software manual.
If any configuration has been
recorder without communication
Extensive software, supplied by order
for the recorder with communication
Ethernet or RS-232/422/485
93
The