12 Specifications

CPR Targets
Category
Compression Depth Target
Compression Release Target
Force
Compression Rate Target
CPRmeter [REF 801-001xx]
The CPRmeter meets the performance requirements of IEC 60601-1, 2nd
and 3rd edition.
Category
Dimensions
Weight
Battery
microSD™ Memory Card
Operating temperature
Storage temperature
Operating Relative humidity
Storage Relative humidity
Operating/storage atmospheric
pressure
IP protection class per ISO/IEC 60529
Specification
> 50 mm
( 2")
Depth accuracy: ±10%
< 2,5 kg (5,5 lbs).
Force accuracy: +1,5 kg, -2,0 kg
(+3,3 lbs, - 4,4 lbs)
100 to 120/min ± 3/min
Specification
154 mm x 64 mm x 28 mm
(6.1" x 2.5" x 1.1")
227g (7,3 oz)
3V Lithium 123 (Li/MnO
The CPRmeter has been tested
with Energizer
®
only Energizer
, Panasonic
®
Duracell
batteries.
®
The CPRmeter has been tested
with 2Gb microSD™ memory
cards. Always test a new card
before using it in data collection.
0° to 40° C (32° to 104° F)
(Fault condition: If the battery has a
short circuit at maximum ambient
temperature, the unit surface
temperature may reach 60° C (140°
F). In this condition the unit will be
inoperable.)
-20° to 70° C (-4° to 158° F)
5% to 95%
5% to 75%
1014 to 572 hPa (101 to 57 kPa)
IP 55