Use And Maintenance - Laerdal CPRmeter Mode D'emploi

Masquer les pouces Voir aussi pour CPRmeter:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Reusable Silicone Cover Kit
The CPRmeter can be fitted with the accessory Reusable Silicone Cover and
the large patient adhesive (single-use). This increases the contact area to the
patient.
Applying the Reusable Silicone Cover and large patient adhesive
1. Check that the silicone cover has been cleaned
and disinfected after last use.
2. Insert the CPRmeter
into the sleeve.
3. Check the label
on the CPRmeter large
patient adhesives resealable
pouch to be sure that they are within their expiration date.
4. Open the package of CPRmeter large patient adhesives and peel one
from the white liner strip to expose the adhesive surface on its underside.
5. Align the bottom of the patient adhesive with the yellow
patient adhesive mount area on the Rear Cover of the
CPRmeter and the adjacent Silicone Cover's bottom surface.
Make sure the channel on the adhesive is directly over the vent
membrane. Press the large patient adhesive into place.
6. Do not peel off the green liner from the attached large patient adhesive
until you are ready to apply the CPRmeter to a patient in an emergency.
Return remaining patient adhesives to their resealable pouch.
IMPORTANT NOTE: Do not delay CPR. If the Reusable Silicone Cover is
generally used with the CPRmeter, the cover and the Large Patient
Adhesives should be applied to the CPRmeter in advance.

Use and maintenance

The CPRmeter use or maintenance is not changed by use of the Silicone
Cover. Clean the Silicone Cover separately from the CPRmeter. Disinfect the
Silicone Cover using the following method:
1. Clean and scrub the cover with a small soft brush, using a mild detergent
until the cover is visibly clean.
2. Rinse in lukewarm water.
3. Submerge the cover in a 0.55% solution of ortho-phthalaldehyde in
accordance with manufacturer's specifications.
4. Rinse again and dry.
The Silicone Cover can also be autoclaved in distilled water at 136 ºC
(210 ºF) and 2.0 kg/cm² for 10-20 minutes.
WARNING: The CPRmeter with the Silicone Cover can not be
considered as disinfected or sterilized.
22