KL 1500 HAL / KL 1500 H - Operating instructions

2. Operation

2.1 Light guide connection

Start by loosening the locking screw (3) of the light guide socket (2) by turning it anti-clockwise.
Insert the light guide as far as it will go and tighten the locking screw by turning it clockwise.
Attention: When inserting light guides with a locating pin, care must be taken to ensure that the pin
fits into one of the two guide slots.
!
Safety information:
ATTENTION!
Light guides can become hot during prolonged use. Please make sure that the light guide is cooled
down before replacing it. It is recommended to turn off the light source for about 5 minutes before
replacing the light guide. A warning sign has been attached
to remind you of this necessity.
The light guide socket (2) is equipped
in addition
with a sensor system (automatic switch off) which
prevents the operation of the lamp unless a recommended light guide is seated in the designated
opening. This protects the user from being blinded during operation without light guide, e.g. when
the light guide is exchanged
by mistake
before the light source is turned off at the on/off switch (1).
If the light guide is removed from the light guide socket (2) during operation, the lamp will switch off.
It will turn back on as soon as the light guide is reinserted into the light guide socket (2). This will not
affect the brightness level set at the light intensity setting (4).
As long as there is no light guide in the light guide socket (2) when the light source is switched on,
the LCD display (5) will show the error message "No Light Guide".

2.2 Power connection:

Insert the three pin power cord of type H05VV-F3G0,75
(optional accessory) into the designated plug in socket (10)
on the back of the instrument. The instrument needs to be
connected to the mains supply (100-240 V AC, 50-60 Hz).
Please ensure that you operate your KL 1500 HAL cold light
source only within the specified voltage range.
When replacing the power cord the same type must be
used.