6.10 ANIMAL (D
The scale can be set to animal (dynamic) weighing for weighing any items that are
unstable or may move.
The scale will use a special filter to minimise the effects of any movement while the
moving animal or unstable samples are on the scale.
Press [Func/Set] and scroll through the list of functions using the ⭡ and ⭣
directional keys
Press [Tare/↵] to enter animal weighing. The display will show "LOAD" and the
Animal/Dynamic weighing symbol
unstable animal or sample on a weighing scale platform.
To use the Animal Weighing function it is necessary to set the amount of filtering
required for the item to be weighed. More active animals will require a higher level
of filtering to give a stable result. Press the [Disp/Mode]. The display will show "Flt
x" where x is a value from 1 to 5. The higher the value the greater the amount of
filter will be.
[Tare/↵] key to accept it.
6.10.1 Animal weighing procedure
With the weighing scale platform empty the indicator display will show
Place containers or blankets onto the platform and press the [0/Esc] key to to
remove the weight of the containers or blankets, alternatively, a long press on the
[Tare/↵] key, will retain the container or blanket value as "NET".
Place the animal or sample to be weighed on the platform.
Press [Tare/↵] key to start to the measurement. The display will show the live
measurements until a stable weight is determined. The time it takes for the stable
value will depend upon the setting of the filter parameter in the first step.
When a stable reading is found, the display will show this value, and the display will
be locked until the [0/Esc] key is pressed. The display will show the "Hold"
symbol while the display is locked.
) WEIGHING
YNAMIC
To increment the value shown press the [⭡] key then press the
The scale is now ready to weigh an
.
"LOAD"