Rohde & Schwarz ESIB7 Manuel D'utilisation page 27

Récepteur de mesure des perturbations électromagnétiques
Table des Matières

Publicité

:[SENSe<1|2>:]DETector:RECeiver[:FUNCtion] POSitive | NEGative | RMS | AVERage | CAVerage |
This command switches on the detectors for single measurements.
Example:
Features:
Mode:
The trace is not selectable; up to four detectors may be switched on simultaneously.
The RMS, NEgative and ACVideo detector cannot be switched on simultaneously.
The AVERage and CAVerage detector cannot be switched on simultaneously.
Selection ACVideo is available only if the instrument is equipped with the linear video output (option
ESIB-B1).
:[SENSe<1|2>:]DETector<1 to 4>:FMEasurement
This command selects the detector for the final measurement (the detector used for the subsequent
final measurement).
Example:
Features:
Mode:
:SYSTem:FIRMware:UPDate <string>
This command starts a firmware update using the files in the set directory.
Example :
Features :
Mode:
This command is an event and has therefore no query and no *RST value assigned.
:TRACe[:DATA]
TRACE1| TRACE2| TRACE3| TRACE4| SINGle | PHOLd | SCAN | STATus | FINAL1
| FINAL2 | FINAL3 | FINAL4, <block> | <numeric_value>
This command transfers trace data from the controller to the instrument, the query reads trace data
out of the instrument.
Receiver
PHOLd yields the level value of the maxhold marker of the bargraph.
1088.7490.01
":DET:REC POS,AVER,QPE"
*RST value:
POS
SCPI:
device-specific
R
"DET:FME POS"
*RST value:
Trace 1, 3 POS
Trace 2, 4 AVERage
SCPI:
device specific
R
":SYST:FIRM:UPD 'C:\V4.32'"
*RST value:
SCPI : conforming
A, VA, BTS, MS
QPEak| ACVideo
NEGative | POSitive | RMS | AVERage |
CAVerage | QPEak | ACVideo
I
E-10