:[SENSe<1|2>:]AVERage:TYPE MAXimum | MINimum | SCALar | VIDeo | LINear
This command selects the trace averaging method.
VIDeo
LINear
Example:
Features:
Mode:
Note:
It is also possible to select the evaluation mode (MAXimum, MINimum, SCAlar) for the
trace with this command. However, it is recommended to use command
DISPlay[:WINDow<1|2>]:TRACe<1...4>:MODE for this purpose. The command
AVERage:TYPE should be used only to select the trace averaging method. Also, the
query reads out the trace averaging mode only.
The following functions are defined but should not be used:
MAXimum (MAX HOLD):
MINimum (MIN HOLD):
SCALar (AVERAGE):
:[SENSe<1|2>:]DETector<1 to 4>[:FUNCtion] APEak | NEGative | POSitive| SAMPle | RMS |
This command switches the detector for recording of the measured value.
Example:
Features:
Modes:
In scan mode of the receiver, the detectors POSitive, RMS, AVERage, CAVerage, QPEak and
ACVideo are available (ACVideo only with option ESIB-B1).
In the analyzer mode, the detectors APEak, POSitive, NEGative, RMS, SAMPLe and AVERage are
available. The value "APEak" (AutoPeak) displays both the positive peak value and the negative
peak value when noise is present. The positive peak value is displayed when one signal is present.
The trace is selected by means of the numeric suffix after DETector.
1088.7490.01
Averaging of logarithmic level values.
Averaging of linear power values prior to their conversion into level values.
":AVER:TYPE LIN"
*RST value:
VIDeo
SCPI:
device-specific
A, VA ("VIDeo" and "LINear" are not available in VA mode)
":DET POS"
*RST value:
Trace1: POSitive
Trace 2: AVERage
SCPI:
conforming
R, A
AVG(n) = MAX(X 1 ...X n )
AVG(n) = MIN(X 1 ...X n )
n
1
×
å
AVG n
( )=
xi
n
=
i
1
AVERage | CAVerage | QPEak | ACVideo
H
E-10