Control panel
4. Connect the transparent tubes of the VADOplex pads to the
connecting tubes of the VADOplex device
5. Switch on the device with the power switch at the front and
wait until the following is displayed:
6. Press the start/stop button to start the therapy.
By pressing the start/stop button during therapy, you can
interrupt the procedure for a break. Pressing the button again
reactivates the device.
Set mode via
With the
button, you can choose from 3 basic modes:
Program 1
Foot
20 Sec.
Program 2
Calf/Foot for
application in the
case of arterial
circulatory
disorders
50 Sec.
Program 3
Hand
20 Sec.
The factory set standard mode is 130 mmHg for 1 second.
Always chose the modes by pressing the
If you want to delete your settings, simply switch off the device with
the power switch and then switch it back on or press the
*Note: If 3 seconds application time is set up for both sides, the interval between
the impulses will raise automatically from 20 seconds to 50 seconds!
(due to slower venous refill time with arterial diseases)
Individual pressure regulation with
button only!
again.
Instruction for use • Footpad
Apply a stockinette or stockings to
prevent skin irritations. Avoid wrinkles.
Place foot as shown by graphic in
the Pad. Note Left and Right side.
The inflation chamber must be
directly under the arch of the foot.
Wrap the inside of the sleeve over the
top of the foot. Cross the sides of the
sleeve around the bridge of the foot
as well and tighten them. Secure with
the Velcro strap.
The connecting tube must be on the
inside of the foot.
Guide the heel strap around the heel
and secure with the Velcro strap at the
outside front. Wrap the Velcro strap
around the sole of the foot.
The sleeves (Pads) are for single patient
use only. Using them for more than one
patient or refurbishing them can cause
spreading infection chains and wound
infections.
The function of the sleeves could be
reduced and limit the therapeutic effect.
CAUTION: Once the treatment phase is complete, used
pads should be disposed in household waste. Do not
send any used pads when returning the controller.
Tubes should be included by the controller.
Single use only!
35