Contraindications; Alarms/Trouble Shooting; Routine Maintenance - OPED VADOplex Professional 1.0 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

www.oped.de | www.oped.ch
Indications
Intended use
Device for improvement of the venous return and arterial flow
with intermittent, pneumatic impulse compression on the upper
and lower extremity. It is equally suitable for specialist staff in
hospitals, with the patient or his relatives using the system
themselves.
Indication
For quick drainage of acute
edema of the upper and lower
extremities (e. g. post traumatic
and postoperative)
Especially suitable for fractures in
the area of the ankle and wrist
After compartment syndrome
Shoulder injuries and surgery
For quick drainage of chronic
edema of the legs and arms
(e. g. chronic lymphedema,
chronic venous insufficiency)
For compartment prophylaxis in
case of serious soft tissue damage
Prevention of deep vein thom-
bosis; when used solely (e. g. in
case of HIT or heparin intolerance)
Prevention of deep vein throm-
bosis; as additional measure to
anticoagulant drugs or for low
risk patients
Chronic recurrent ulcera cruris
(arterial, venous, mixed)
Venous congestion
Diabetic foot syndrome / ulcers
Peripheral artery occlusive disease
(PAOD) and arterial circulatory
disorders of the extremities
Recommended
application time
post traumatic: continuous
(with breaks if required)
postoperative: 6 to 8 hours daily
(with breaks if required)
At least 4 hours per day
with breaks if required
over several weeks
Continuous, without breaks
At least 11 hours per day, gradu-
ally reducing the application time
until full load is possible
At least 6 hours per day in
the high-risk phase, gradually
reducing the application time
until full load is possible
At least 4 hours per day
with breaks if required
over several weeks
At least 4 hours per day
with breaks if required
over several weeks
At least 4 hours per day,
with breaks if required
over several weeks

Contraindications

For the following types of patients, risks and benefits have to be
assessed:
Patients whose heart could be impaired by an increased reverse
current, patients with decompensated cardiac insufficiency, acute
deep leg vein thrombosis or pulmonary embolism as well as throm-
bophlebitis or massive infections of the extremities treated. If you
have any question, please ask your doctor.

Alarms/trouble shooting

Error code 1
Air tube disconnected? / exchange pad
Error code 2
Air tube bent?
Error code 3
Pressure too low – exchange pad
Error code 4
Pressure too high – exchange pad

Routine Maintenance

Interval
Routine Maintenance
As required
Thoroughly clean external surfaces of the
controller before transfer between patients
and after prolonged periods of storage.
Attention should be paid to the fan grills,
ensure these are clean at all times.
Annually/
Inspect the AC powersupply plug and cable
3.000 hours
for damage. If this display appears, please
send back the VADOplex Foot Pump System
to the manufacturer for service.