Immunity test
Conducted disturbances
induced by RF fields
IEC 61000-4-6
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Note: UT is the AC mains voltage prior to application of the test level. If in the unlikely event the Philips
Avent single/double electric breast pump Advanced is locked inoperative in a function by
disturbances from other devices beyond the levels stated in the table above, the device will shut-off
completely and restart. In that case make sure the distance to the other device causing this
disturbance is enlarged.

Explanation of symbols

The warning signs and symbols are essential to ensure that you use this device safely and correctly
and to protect you and others from injury. Below you find the meaning of the warning signs and
symbols on the label and in the user manual.
Symbol
Explanation
Indicates to follow the instructions for use.
Indicates important information such as warnings and cautions.
Indicates usage tips, additional information or a note.
Indicates the manufacturer.
Indicates the manufacturing date.
Immunity test level
3 Vrms, before
modulation is applied
0.15 MHz – 80 MHz
6 Vrms, before
modulation is applied, in
ISM and amateur radio
bands between 0.15 MHz
and 80 MHz
80 % AM at 1 kHz
0 % UT; 0.5 cycle at 0°,
45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle and 70 %
UT; 25/30 cycles single
phase at 0°
0 % UT; 250/300 cycle
Electromagnetic environment -
guidance
Mains power quality expected of a typical
commercial or residential environment. At
power interruptions the breast pump will
switch off.