GROUNDING INSTRUCTIONS

The plug must be plugged into an outlet that is properly installed grounded. In the event of an electrical short
circuit, grounding reduces the risk of electric shock .
The electrical outlet, which must be a 220V-240V 50Hz, and at least a 10-Amp grounded outlet,
should be located on the wall within reach of the unit's power cord.
A Ground Fault Interrupter residual current device (RCD) should be installed on the circuit supplying power to
the unit. According to each local regulation.
DANGER
- Improper use of the grounding plug can result in a risk of electric shock. If repair
or replacement of cord or plug should be done by authorized personel. Make sure grounding is correct.
Check with a qualified electrician or serviceman if the grounding instructions are not completely
understood, or if in doubt as to whether the product is properly grounded.
Do not use extension cable.