GUIDANCE AND MANUFACTURER'S
DECLARATION ELECTROMAGNETIC IMMUNITY
Mesh–nebulizer (membrane) AGU N7 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the mesh–nebulizer (membrane) AGU N7
should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000–4–2
Electrical fast
transient / burst
IEC 61000–4–4
Surge
IEC 61000–4–5
Voltage Dips,
short interrup-
tions and volt-
age variations
on power supply
input lines
IEC 61000–4–11
Power frequen-
cy (50, 60 Hz)
magnetic field
IEC 61000–4–8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
64
IEC 60601
test level
+ 6 kV contact
+ 8 kV air
+ 2 kV for power
supply lines
+ 1 kV for
input / output lines
+ 1 kV line(s)
to line(s)
+ 2 kV line(s)
to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec.
3 A / m
Compliance level
+ 6 kV contact
+ 8 kV air
+ 2 kV for power
supply lines
Not applicable
+ 1 kV differential
mode
Not applicable
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec.
3 A / m
Electromagnetic
environment–guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material,
the relative humidity
should be at least 30 %
Main power quality
should be that
of a typical commercial
or hospital environment
Main power quality
should be that of a typical
commercial or hospital
environment
Main power quality
should be that of a typical
commercial or hospital
environment. If the user
of the AGU N7 requires
continued operation
during power mains
interruptions, it is recom-
mended that the AGU N7
be powered from an
uninterruptible power
supply or a battery
The AGU N7 power
frequency magnetic
fields should be at levels
characteristic of a typi-
cal location in a typical
commercial or hospital
environment