TROUBLESHOOTING

If any trouble occurs while you are using the device, please check the following list first.
Problem
Low atomization
Non–operation even
if power switch is on
(power indicator does
not light and no mist
comes out)
No atomization even if
power switch
and indicator are on
Power indicator turns
into orange light
constantly
Power indicator turns
into orange light and
flashing
WARNINGS
If the device does not nebulize normally after taking the above–mentioned procedure,
please contact the store where you purchased the device or the nearest dealer.
Never operate this device if any of the parts are not working properly or have been
damaged.
We recommend to exchange the medication cup after approximately 1 year,
and 2 years for the device.
60
Possible Cause
Low battery power
The stains on the electrode
cause a fault connection
The mesh holes are clogged
The mesh is broken
The batteries are inserted
in the wrong direction
Batteries are dead
Fault connection between
adapter and main unit
The medication cup is not
installed properly
The mesh holes are clogged
The mesh is broken
Low battery power
Fault connection of electrode
and main unit
The mesh is broken
Action
Replace the batteries or use
the adapter
Use rubbing alcohol to clean
electrodes
Refer to Cleaning and Disin-
fection procedure to clean the
medication cup
Replace the medication cup
Follow (+) and (–) mark on the
battery cover to re–insert the
batteries in correct direction
Replace the batteries or use
the adapter
Check and reconnect
the adapter to main unit
Refer to Set up the medication
cup procedure to re–install
the medication cup
Refer to Cleaning and
Disinfection procedure
to clean the medication cup
Replace the medication cup
Replace the batteries or use
the adapter
Use rubbing alcohol to clean
electrodes and re–stall the
medication cup
Replace the medication cup