TROUBLESHOOTING

In case of malfunctions in the device operation, refer to the following table for its
elimination.
Malfunctions
Possible reasons
Extremely
The medication chamber is not well
low nebulizing
installed
speed
The drugs do not enter the
membrane for more than 15
seconds
The membrane of the medication
chamber is clogged
The electrodes on the medication
chamber are littered with medicine
or water
The electrodes on the nebulizer and
medication chamber are contami-
nated
After turning
The medication chamber is not well
on, the power
installed
indicator lights
up for 15 sec-
The medication chamber is empty Fill in the medication chamber.
onds and then
turns off imme-
diately
The medicine does not enter the
membrane
The electrodes on the nebulizer and
medication chamber are contami-
nated
The power indi-
The batteries are installed incor-
cator is off and
rectly
the nebulizer
does not work
Low battery
The micro - USB cable is not prop-
erly connected to the nebulizer
126
Troubleshooting
Install the medication chamber
correctly and start nebulizing.
(See page 108)
Place the nebulizer at such an
right angle that the drugs get on
flow to the membrane
Clean the medication chamber. If
this does not solve the problem,
replace a new chamber
Clean the electrodes and start
nebulizing again. (See page 118)
Clean the electrodes and restart
the power. (See page 118)
Install the medication chamber
correctly and start nebulizing.
(See page 101)
(See page 113)
Place the nebulizer at such angle
that the can get on the mem-
brane
Clean the electrodes and restart
the power. (See page 118)
Reinstall the batteries, observing
the polarity and restart the device.
(See page 110)
Replace the batteries and restart
the device. (See page 110)
Connect it correctly and restart
the device. (See page 110)