REPLACEMENT OF THE MEDICATION CHAMBER
The service life of a medication chamber is 12 months under normal conditions
(when used three times a day for a total use of 30 minutes). However, spray-
ing efficiency may begin to deteriorate in less than 12 months depending on how
you use it or when using certain types of medications. If the nebulizer does not
work correctly or the spraying speed is significantly reduced after cleaning, you
should replace the medication chamber with a new one. (If you want to purchase
a medication chamber, contact the authorized seller from whom you purchased
the product or the nearest distributor).
1. Remove the medication chamber from the nebulizer.
Open/dissemble the main unit. Press the PUSH button on the back of the main
unit and pull the container towards the front of the main unit.
WARNING
In order to avoid the device damage, make sure that you press the PUSH
button before removing the container from the main unit.
In order to avoid the membrane rupture, do not press on it or poke it with
your finger or other objects.
Magnetic
lock
1.1
a) Take off the cover by pulling a small flange to take a cover away. (1.1).
b) Press the PUSH button to remove the medication chamber from the main unit (1.2).
122
A small flange
to take cover away
Press to release
medication chamber
1.2