GUIDANCE AND MANUFACTURER'S
DECLARATION ELECTROMAGNETIC IMMUNITY
The Mesh–nebulizer (membrane) AGU N7 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the Mesh–nebulizer (membrane) AGU N7
should assure that is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000–4–6
150 KHz to
80 MHz
Radiated RF
20 V / m
IEC 61000–4–3
80 MHz
to 2,5 GHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular / cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured field strength in the location in which the Mesh–nebulizer (membrane)
AGU N7 is used exceeds the applicable RF compliance level above, the Mesh–nebulizer
(membrane) AGU N7 should be observed to verify normal operation. If abnormal perfor-
mance is observed, additional measures may be necessary, such as re–orienting or relo-
cating the Mesh–nebulizer (membrane) AGU N7
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V / m.
EN
Compli-
Electromagnetic environment–guidance
ance
level
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any
part of the AGU N7 including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance:
d = 1,2 √P
20 V / m
d = 1,2 √P 80 MHz to 800 MHz
d = 2,3 √P 800 MHz to 2,5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in
each frequency range
Interference may occur in the
vicinity of equipment marked
with the following symbol:
.
b
,
a
65