Recommended separation distances between portable and mobile RF communications equipment
and the device – for device that is not LIFE-SUPPORTING.
Recommended separation distances between portable and mobile RF communications equipment and the
device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the device
as recommended below, according to the maximum output power of the communications equipment
Rated maximum
output power of
150 kHz to 80 MHz
transmitter
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the trans-
mitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
15

MANUFACTURER'S WARRANTY

The warranty period for the main unit is 24 months, for the medication chamber – 12 months from the
date of purchase. The warranty is only valid on presentation of the sales receipt and the warranty card
completed by the authorized dealer. The warranty does not apply to consumables, constituent parts
and components, such as: protective cover, batteries, tip/mouthpiece, carrying/storage case, micro-
USB-C cable, inhalation masks (S, M), package.
Warranty and free maintenance service is not provided in case of:
• violation of operating instructions;
• damage caused by intentional or erroneous actions of the consumer due to improper or negli-
gent handling;
• evidence of mechanical impact, dents, cracks, chips, etc. on the nebulizer housing, evidence
of opening the device, disassembly, unauthorized repair, ingress of moisture, exposure to
EN
Separation distance according to frequency of transmitter (m)
80 kHz to 800 MHz
d = 1,2×√P
d = 1,2×√P
0,12
0,38
1,2
3,8
12
0,12
0,38
1,2
3,8
12
800 kHz to 2,5 GHz
d = 2,3×√P
0,23
0,73
2,3
7,3
23
79