Diode Testing - Iso-Tech IDM 97 Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3.6 Diode testing

1. Turn the function switch to "
2. Plug the black lead into the "COM" input terminal and the red lead into the "V -
3. Connect the probes to the diode to be tested. With a non-faulty silicon diode, the voltage shown in the forward direction
will be between 0.400 and 0.900V. When the diode being tested is faulty, "000" (short-circuit) or OL (open circuit) will be
shown.
4. When tested in the opposite direction (blocking direction), "000" or a different value will appear for a faulty diode.
".
è
20
- " input terminal.

Publicité