Analogschnittstelle; Schalter; Analog Interface; Interface Analogique - Ismatec ISM 915 Mode D'emploi

Masquer les pouces Voir aussi pour ISM 915:
Table des Matières

Publicité

speed IN
direction
input 1
start
+26 V
remote
DC
input 2
GND
output 2
speed OUT
output 1
+5 V
speed intern
DC
Digitale Eingänge (TTL-Pegel)
Digital inputs (TTL-level)
Entrées numériques (niveau TTL)
Pin 2, remote
Pin 3, start
Pin 4, direction
Pin 6, input 1
Pin 8, input 2
Pin 13, speed intern
Analoge Eingänge
Analog inputs
Entrées analogiques
Pin 5, speed IN
0–5 V
/ 0–10 V
DC
DC
0–20 mA / 4–20 mA
Universal Ausgänge (PWM)
Universal outputs (PWM)
Sorties universelles (PWM)
Pin 14, output 1
Pin 15, output 2
Analog-Ausgang
Analog output
Sortie analogique
Pin 9, speed OUT
0-10 V
/ 0-7.2 kHz
DC
MCP PROCESS, 14-014, REV. C

Analogschnittstelle

Pin 1, GND (Masse)
Bezugspotential für alle anderen
Eingänge.
Pin 2, remote
Für Umschaltung zwischen manueller
Bedienung und der Analogschnittstelle.
Zur Aktivierung der Analog-Schnittstelle
muss Pin 2 mit Pin 1 (GND)
verbunden werden.
Pin 3, start
– Im Normalbetrieb (Pin 2 offen) dient
Pin 3 als Fußschalter
(Einstellungen siehe Seite 17)
– Im Remote-Betrieb (Pin 2 auf GND)
startet die Pumpe bei Verbindung
mit Pin 1 (GND)
Pin 4, direction
Wenn offen, dreht die Pumpe
im Uhrzeigersinn; wenn mit Pin
1 (GND) verbunden, dreht sie im
Gegenuhrzeigersinn
Pin 5, speed IN
Externe Drehzahlsteuerung
(0 – 5V, 0 – 10V, 0 – 20mA, 4 – 20mA)
Eingangsimpedanz und Wahl-
möglichkeiten mittels DIP-Switch im
Geräteinnern (siehe Seite 38)
Pin 6, input 1
Pin 8, input 2
Digitale Eingänge (TTL-Pegel). Aktiv, wenn
mit Pin 1 verbunden. Können nur über
RS232-Schnittstelle abgefragt werden; z.B.
mit einer Software wie ProgEdit.
Pin 7, +26V
DC
Es stehen ca. +26 V
zur Verfügung
DC
(max. Strom 1A).

Analog interface

Pin 1, GND (ground)
Reference potential for all other inputs
Pin 2, remote
For changing between manual control
and analog interface. For activating
the analog interface, pin 2 must be
connected with pin 1 (GND).
Pin 3, start
– In normal operation (pin 2 open)
pin 3 operates the footswitch
(settings see Page 17)
– In remote operation (pin 2 to
GND) the pump starts when
connected to pin 1 (GND)
Pin 4, direction
In the open position the pump turns
clockwise; when connected to pin 1
(GND) it turns counter-clockwise.
Pin 5, speed IN
External speed control
(0 – 5V, 0 – 10V, 0 – 20mA, 4 – 20mA)
Input impedance and input range can
be selected via a dip-switch inside the
pump (see Page 38)
Pin 6, input 1
Pin 8, input 2
Digital inputs (TTL-level). Active when
connected to pin 1. Are only retrievable
via RS232 interface; e.g. with a software
like the ProgEdit.
Pin 7, +26V
DC
About +26 V
are available
DC
(max. current 1A).

Interface analogique

Pin 1, GND (masse)
Potentiel de référence pour toutes les
autres entrées
Pin 2, remote
Pour commuter du service manuel à
l'interface analogique. Pour activer
l'interface analo-gique, le pin 2 doit être
connecté au pin 1 (GND).
Pin 3, start
– En exploitation normale (pin 2
ouvert), le pin 3 sert d'interrupteur au
pied (selection voir en Page 17)
– En exploitation à distance (pin 2 sur
GND), la pompe se met en route dès
qu'elle est connectée au pin 1 (GND)
Pin 4, direction
Si ouvert, le sens de rotation de la
pompe est celui des aiguilles d'une
montre; si relié avec le pin 1 (GND),
elle tourne dans le sens contraire des
aiguilles d'une montre.
Pin 5, speed IN
Réglage externe du nombre de tours
(0 – 5V, 0 – 10V, 0 – 20mA, 4 – 20mA)
Impédance d'entrée et réglage de
zone au moyen de l'interrupteur DIP à
l'intérieur de l'appareil (voir Page 38)
Pin 6, input 1
Pin 8, input 2
Entrées numériques (niveau TTL). Actifs
si reliés au pin 1. Ne peuvent être
interrogés que par l'interface RS232,
p.e. avec le logiciel ProgEdit.
Pin 7, +26V
CC
Environ +26 V
sont à disposition
CC
(courant maximal 1A)