CURING LIGHT OPERATING STATES
Plus
There are three distinct operational states when operating the wand:
I.
Idle: The unit is Idle when it is not Curing or in Sleep mode and the battery
is CHARGED or LOW. The user may cycle between 5 second, 10 second, and
20 second curing modes by pressing the mode switch. Curing cycles are also
initiated when the unit is in the Idle state.
II.
Curing: A Curing cycle is initiated by pressing, and releasing, the trigger switch
while the unit is Idle. A 5 second, 10 second, or 20 second cycle will be started,
depending on the status indicated by the Curing Mode LEDs. Once a Curing cycle
is started, the Curing Light LED will be turned on and a tone sounds. Depending
upon which mode was selected, a tone will sound every five seconds until the
cycle is complete, at which time a final, longer duration, tone sounds.
Note: Pressing the Trigger Switch for a second time during a Curing cycle
immediately interrupts the Curing cycle and causes the final tone to sound,
regardless of how much time has elapsed since beginning the cycle (pressing the
Mode Selection Switch during a Curing cycle has no effect).
III. Sleep: The Demi
goes to Sleep mode to conserve battery life after
Plus
approximately 8 minutes of no activity. It can be awakened by pressing either the
mode or trigger switch once, at which time the unit will be in the Idle state. The
unit will reflect the last cure mode chosen. The mode or trigger switches can then
be pressed to switch between modes or to initiate a cure. While in Sleep mode,
all LEDs are turned off and the curing light goes into a low power state. Battery
life (if left uncharged and unused after receiving a full charge) is approximately 6
months whether the battery is in the handpiece or not.
OPERATE THE DEMI
CURING WAND
Plus
I.
Select the desired curing mode by pressing the Mode Selection Switch (Figure 9, 1.)
II.
Place the Light Guide (2.) over the tooth, as close to the tooth as possible without
touching it. Press and release the Trigger Switch (3.). The unit will sound a tone at
5 second intervals as previously described.
MAINTENANCE:
CLEANING / DISINFECTING / STERILIZING
I.
Fiber-Optic Light Guide:
a)
Immerse/thoroughly drench light guide in a surface decontaminate cleaner
(soap and water cleaning solution).
b)
After 10 minutes, scrub with an instrument cleaning brush until visibly free
of debris.
c)
Rinse all parts with warm water.
II.
Fiber-Optic Light Guide - Sterilizing:
a)
Sterilize in a steam autoclave using distilled water at a temperature of
270°F (132°C) for 20 minutes.
b)
Do not add chemical solutions to the autoclave.
c)
Do not use a chemical autoclave.
III. Optics Maintenance Kit (PN 21042) – After numerous autoclave cycles, a brown
or gray discoloration can accumulate on the polished ends of the light guide.
This is a residue that forms during the autoclave drying cycle. It is recommended
that the Optics Maintenance Kit be used every 50 autoclave cycles. Following the
instructions included in the kit helps promote optimum curing performance.
To remove composite that adheres to the tip of the light guide:
·
Soak gauze in alcohol.
·
Wrap the gauze around a razor blade.
·
Rub the tip of the light guide with the gauze-protected razor blade until
the composite is removed.
Note: Never use a blunt instrument to remove composite because it will
scratch the glass.
IV.
Handpiece and Battery Charger – Plastic parts can be cleaned following these
guidelines:
a)
All plastic parts may be wiped with CaviWipes™ following the direction on
the package. A quaternary ammonium compound product like CaviCide™/
CaviWipes™ or similar is recommended (containing 20% alcohol or less).
b)
Do not use: Denatured alcohol, Lysol®, phenol, ammonia complex or iodine
complex solutions.
c)
Wipe, do not spray, solution onto plastic parts. Prevent liquids from
entering openings on unit.
CAUTION: Do not immerse the plastic parts in solutions. Use of solutions other
than those recommended may damage plastic parts and will void product warranty.
This unit is not autoclaveable.
INFECTION CONTROL MEASURES
CAUTION: To prevent cross-contamination, a disposable plastic sleeve must be
used over the Demi
with each use.
Plus
WARRANTY
Kerr hereby warrants that for a period of three years (one year for batteries) from
the date of purchase, this instrument shall be free from defects in material and
workmanship and will perform satisfactorily under normal use and service.
To register your warranty, complete the online warranty registration within 30 days of
purchase at www.kerrdental.com/warranty. If multiple units have been purchased, fill
out one registration form per unit. Please have your dealer invoice available. You can
also register your warranty by calling 1.800.KERR.123 (1.800.537.7123).
THE WARRANTY STATED HEREIN IS THE SOLE WARRANTY APPLICABLE TO KERR
PRODUCTS. KERR EXPRESSLY DISCLAIMS ANY AND ALL OTHER WARRANTIES EXPRESSED
OR IMPLIED, INCLUDING WARRANTIES OR MERCHANTABILITY OR FITNESS FOR A
PARTICULAR USE. KERR'S LIABILITY WITH RESPECT TO ITS PRODUCTS IS EXPRESSLY
LIMITED TO THE REMEDIES SET FORTH ABOVE. THE REMEDIES ARE THE BUYER'S
EXCLUSIVE REMEDIES. KERR SHALL UNDER NO CIRCUMSTANCES BE LIABLE FOR
INCIDENTAL OR CONSEQUENTIAL DAMAGES. THE WARRANTY DOES NOT APPLY
TO ANY DAMAGE OR DEFECT CAUSED BY THE USER'S FAILURE TO THE FOLLOWING
INSTRUCTIONS FOR USE OR THE OTHER USER ERROR IN OPERATING OR MAINTAINING
THE INSTRUMENT. ALTHOUGH KERR OFFERS SERVICES TO REPAIR INSTRUMENTS AND
PROVIDES REPLACEMENT PARTS AFTER THE EXPIRATION OF THE WARRANTY, KERR ONLY
GUARANTEES THE AVAILABILITY OF SUCH SERVICES AND REPLACEMENT PARTS FOR TWO
YEARS AFTER THE INSTRUMENT MODEL HAS BEEN DISCONTINUED.

ACCESSORIES AND REPLACEMENT PARTS

Part No.
Fiber Optic Light Guides
21020
8 mm Curved Turbo Light Guide
921551
11 mm Curved Turbo Light Guide
20812
13 mm Curved Light Guide
Part No.
Eye Protection
20816
Protective Light Shield
Part No.
Accessories
21042
Optics Maintenance Kit
20399
Disposable Hardness Disk, (pack of 6)
910726
Handheld LED radiometer
Part No.
Replacement Parts
921918-1
Replacement Kit, Battery, Demi
921919-1
Replacement Kit, Charger, Demi
921920-1
Replacement Kit, Wand, Demi
921656
Plug-in Universal Power Supply