Connection B (J2) - Logic Output Interconnections

J2 OUT
15
FAULT (+24V)
R 2 (+24V)
ST ART (+24V)
LOW SPEED (+24V)
R 1 (+24V)
SPEED MONITOR (0-10 V)
9
Logic output connector
All the logic output from the controller must be
connected at P2 mating connector.
With the optional P2 mating connector make the
connections with AWG 24 (0.25 mm
the pins indicated in the figure to obtain the desired
capability.
The following table describes the signals available on
the connector.
PIN
DESCRIPTION
1-2
Analog output voltage of DC power drawn by
the turbopump (pin 2 positive, pin 1 negative).
1 Vdc ≡ 10 W
1-9
Analog output voltage (0 - 10 V) of pump
speed (pin 9 positive, pin 1 negative).
4-11
RELAY 1 signal 24 V, 60 mA, optically isolated
output (pin 11 positive, pin 4 negative).
The output voltage will be present when the
rotational speed of the pump is higher than the
selected speed threshold.
5-12
LOW SPEED signal, 24 V, 60 mA, optically
isolated
output
negative).
The output voltage will be present when the
low speed mode is selected, either through the
hand held terminal or the remote signal.
6-13
START signal 24 V, 60 mA, optically isolated
output (pin 13 positive, pin 6 negative).
The output voltage will be present when the
START pushbutton on hand held terminal is
pressed, or the remote start is present, until
NORMAL operation is reached.
8
FAULT
R 2
ST ART
LOW SPEED
R 1
POW ER MONITOR (1 Vdc = 50 W)
POWER AND SPEED MONITOR (COM)
1
2
) or smaller wire to
(pin
12
positive,
pin
PIN
7-14
R2 signal 24V, 60 mA, optically isolated output
(pin 14 positive, pin 7 negative).
The output voltage will be present upon the
programmed condition delay YES or delay NO
(see the cycle diagram in the following pages).
If YES is selected, R2 is off and the
is zero over all run up time, then:
a) If
R2 = OFF
b) If
R2 = ON
If NO is selected:
a)
R2 = OFF
b)
R2 = ON
8-15
FAULT signal 24V, 60 mA, optically isolated
output (pin 15 positive pin 8 negative).
The output voltage will be present when a fault
condition is detected.
The following figure shows a typical logic output
connection (relay coil) but any other device may be
connected e.g. a LED, a computer, etc., and the related
simplified circuit of the controller.
5
53
TECHNICAL INFORMATION
DESCRIPTION
running
speed
>
speed
running
speed
<
speed
If running speed > speed threshold
If running speed < speed threshold
8
8
15
15
9
9
1
1
Typical output connection
87-900-883-01 (C)
output
threshold
threshold
+24 V