GENERAL INFORMATION

This equipment is destined for use by professionals.
The user should read this instruction manual and any
other additional information supplied by Varian before
operating the equipment. Varian will not be held
responsible for any events occurring due to non-
compliance, even partial, with these instructions,
improper use by untrained persons, non-authorised
interference with the equipment or any action contrary
to that provided for by specific national standards.
The Turbo-V 70 Switching PCB controller is a
microprocessor-controlled,
converter
with
self-diagnostic
features.
It incorporates all the facilities required for the
automatic operation of the Turbo-V 70 pump series.
The controller drives (within ten steps) the Turbo-V 70
pump during the starting phase by controlling the
voltage and current respect to the speed reached by
the pump.
Remote start/stop, pump status signals, forepump
start/stop, interlock control (for pressure switch, water
flow switch, etc.) capability, are provided via auxiliary
connectors. The controller needs to be powered by a
power supply between 90 and 265 Vac with a
frequency of 50/60 Hz. An internal controller circuitry
verifies the power supply voltage and generates the
necessary voltages to supply the controller. An hand
held terminal is available as an option to control and
display
the
controller
paragraphs contain all the information necessary to
guarantee the safety of the operator when using the
equipment. Detailed information is supplied in the
appendix "Technical Information".
This manual uses the following standard protocol:
!
The warning messages are for attracting the attention
of the operator to a particular procedure or practice
which, if not followed correctly, could lead to serious
injury.
The
caution
messages
procedures which, if not followed, could cause damage
to the equipment.
The notes contain important information taken from the
text.
solid-state,
frequency
and
self-protection
operation.
The
following
WARNING!
CAUTION!
are
displayed
before
NOTE
STORAGE
When transporting and storing the controllers, the
following
environmental
satisfied:
temperature: from -20 °C to + 70 °C
relative humidity: 0 - 95%
(without conden-sation)

PREPARATION FOR INSTALLATION

The controller is supplied in a special protective
packing. If this shows signs of damage which may have
occurred during transport, contact your local sales
office.
When unpacking the controller, ensure that it is not
dropped or subjected to any form of impact.
Do not dispose of the packing materials in an
unauthorized manner. The material is 100% recyclable
and complies with EEC Directive 85/399.
Before installing the controller you must prepare some
interconnection cables, i.e. the power interconnections,
the logic interconnections and the pump connections.
The following pictures show the interconnections
between pump and controller and between controller
and power transformer.
Detailed
information
transformer chatacteristics (available as an option) are
supplied in the appendix "Technical Information".