Warning Messages; Error Codes - IKA Tube Mill control Mode D'emploi

Masquer les pouces Voir aussi pour Tube Mill control:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 23
Setting "Timer" value (timer sec) in working screen:
 Press "Timer" key (E) to activate the timer setting in
the working screen.
 Turn the "Rotating/Pressing" knob (B) to change the
value.
 Press "Timer" key (E) again to complete the setting.
 Current values are stored.
• Interval function:
When set "Int" value with the "Rotating/Pressing" knob
(B), the value can be set from 5 to 60 seconds where ad-
justment magnitude is 5 second.
Note: If both "Int" function and "Timer" function are
activated, the "Int" value should be less than "Timer"
value. Otherwise, the interval function will not work.
Warning 1:
Warning 2:
Corrective action:
The fault is shown by an error code on the display (C) as following if the error occurs.
Proceed as follows in such cases:
 Switch the device off with mains switch (A).
 Carry out corrective measures.
 Restart the device.
Error code Cause
Error 3
• Inside temperature is too hot
Error 4
• Motor can't running normal (can't
start or not reach the target speed)
• Overload
Error 25
• Electronic hardware failure
Error 26
Error 70
Error 71
Error 86
• Voltage for electronically
commutated motors is too low
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department.
- Send the instrument for repair, including a short description of the fault.

Warning messages

The protective hood is not closed.
The milling tube is not properly adapted.
Adapt the milling tube properly and close the protective hood, then start the test.

Error codes

Setting "Int" timer (int sec) in working screen:
 Press "Int" key (F) to activate the interval setting in the
working screen.
 Turn the "Rotating/Pressing" knob (B) to change the
value.
 Press "Int" key (F) again to complete the setting.
 Current values are stored.
Effect
Solution
motor off
- Switch off the instrument and allow it to cool down.
- Restart the instrument.
motor off
- Switch off the instrument.
- Check and reduce the grinding material.
- Reduce the grinding speed at the beginning.
- Restart the instrument.
motor off
- Send the instrument for repair.
motor off
- Switch off the instrument.
- Check if voltage give on the type plate corresponds
to the mains voltage.
20

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

20000005401

Table des Matières