80 Annexe B : Compatibilité électromagnétique (CEM)
Recommended separation distances between portable and mobile RF communications
equipment and the equipment
Rated max. output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Separation distance according to frequency of transmitter (m)
150 KHz to 800 MHz
d= 1.2
0.1m
0.4m
1.2m
4.0m
12.0m
Welch Allyn® Diagnostic Cardiology Suite Resting ECG
800 MHz to 2.5 GHz
d= 2.3
0.2m
0.7m
2.3m
7.0m
23.0m