7.9 Setting the Motorized Polarizer

Remove your specimen from the stage.
Leica DMI3000 B:
• Set the condenser to the bright fi eld position.
• Insert the analyzer into the analyzer slot on
right side of the stand.
• Activate the polarizer.
• Turn the polarizer until you have the optimal
dark position.
Leica DMI4000 B and Leica DMI6000 B:
For manual condensers, proceed as described
above for the DMI3000 B.
• Select the POL method (one of the variable
function buttons on the stand). If the analyzer
is present on the Fluo turret as an analyzer
block, it will move into position automatically. A
manual analyzer must be positioned by hand.
In the case of motorized condensers with mo-
torized polarizers, the polarizer will move into
position automatically.
• Insert the centering key in the opening provid-
ed on the condenser (Fig. 96).
• Set up optimal darkening. (The analyzer must
be in place.)
• Remove the centering keys.
Replace your specimen on the stage.
7.10

Adjusting the Light Sources

Transmitted light axis (TL) with lamp housing
107/2
The lamp housing 107/2 with a 12V 100W halo gen
lamp is fi xed. Centering the lamp is not required.
Lamp housing 107 L
for 12V 100W halogen lamp
The lamp can be adjusted using the screws (98.2)
and the button (98.3).
• Place a sheet of white paper under the fi eld
diaphragm.
• Adjust the lamp to create an evenly bright spot
on the paper.
Incident light axis (IL) with lamp housing 106 z
• When a supply unit is used, it is turned on fi rst.
• Activate the incident light axis (for Leica
DMI4000/6000 B with the TL/IL function button.
FLUO will appear on the LeicaScreen).
• Insert the lamp adjustment refl ector (Fig. 99) in
the fi lter turret in place of a fi lter cube.
Make a note of the designation of the re placed
fi lter cube.
7. Start-up
77