Error messages

The following error messages may occur (Fig.21) and have the causes
listed below:
Er 1 Sensor not connected or defective (open circuit).
Er 2 Analog output loaded with <50kΩ, output voltage too low.
Er 3 Internal temperature above +76 °C.
Er 5 Sensor not in medium (no change in medium temperature
after heater has been on for 3 minutes).
Er 6 Sensor short circuit.
Er 7 The maximum set HI-TEMP is exceeded.
Er 8 Calibration error (re-calibration by field service required).
Er 9 IC107 (4066) defective
Er 10 T_105 defective, OUT_TEMP >480 °C (4.8V)
Er 11 Supply voltage <7.5V
Er 12 Supply voltage >16.1V

Analog output

The actual temperature is represented as a voltage (1V corresponds to
100°C) at the 3.5mm jack of the twin-jack adapter.
Tolerance:
0...100 °C ± 10 mV
100...400 °C ± 25 mV
The internal resistance of the connected measuring instrument or recorder
must be greater than 50 kΩ.
12

Accessories

H 36
Retaining bar
H 44
Retaining clip
H 50
Twin-jack adapter
H 52
Power supply kit
H 60
Short sensor
60 mm stainless steel
H 62
Standard sensor 270 mm stainless steel
H 64
Glass sensor
50 mm glass
H 66
Glass sensor
270 mm glass
H 70
Extension cable, 1 m long
H 71
Extension cable, 2.5 m long
Ak 2.5
Analog cable

Technical data

Measuring range:
°C
Resolution:
K
K
Measuring accuracy:
K
Setting accuracy:
K
Control deviation:
K
Supply voltage:
V DC
Current consumption:
mA
Ambient temperature:
°C
Relative humidity:
%
ON time:
%
Degree of protection:
Cable length:
mm
Plug:
Connection:
Dimensions: (W x D x H)
mm
Weight:
kg
-10 ... 400
0.1 (-10 ... +100 °C)
1 (100 ... 400 °C)
±0.3 + sensor tolerance Pt 1000
DIN IEC 751 Class B
1
± 1
8 ... 16
16 (at 9V)
0 ... 60
80
100
IP 44
350
6-pin DIN 45 322
DIN 12 878 Class 2
70 x 150 x 65 (without sensor)
0.2
ETSD 0105

Publicité