Acoustic Warning Signal; Monitoring Functions - B.Braun Aesculap Lektrafuse GN200 Mode D'emploi/Description Technique

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 42
Aesculap
Lektrafuse HF generator GN200
2.5

Acoustic warning signal

Status
System failure
REGRASP error
System error
System warning
Start HF discharge
HF discharge active
End of HF discharge
Power up self-test
2.6

Monitoring functions

Self-test
As soon as it is switched on, the unit performs a self-test of the control
elements, the acoustic warning signal, the microprocessor and the hard-
ware function. During this phase, the message "Performing Self-Test"
appears in the display 5.
Fig. 1
Continuous test cycle during operation
During operation, safety-relevant functions and signals are monitored
through a continuous test cycle. As soon as a critical error is detected, the
HF generator terminates the HF activation. In display 5, the error number
of the corresponding error is displayed and an acoustic warning signal is
emitted, see Troubleshooting list.
6
®
Acoustic warning signal
3 repetitions of: 1 long beep, 3 KHz, 1 beep, 2 KHz Following each system failure, if identified (F001,...)
3 repetitions of: 1 beep, 2 KHz,
1 beep, 3 KHz
3 beeps, 3 KHz
3 beeps, 3 KHz
2 beeps, 480 Hz
1 beep, 480 Hz
In Standard mode: Repetition with 1.8 Hz
In Plus mode: Repetition with 2.8 Hz
3 beeps, 530 Hz
1 beep, 2 KHz
Notes
Following each regrasp error, if identified
Following each failure report, if identified (E001,...)
Following each warning, if identified
When starting the HF discharge
Continually during HF discharge
When a sealing process has been successfully completed (no
failure reports or errors)
During the self-test, in order to check that the signal tone is
functioning