Technical Information
Examples

Window Meanings

Tab. 15
Accessories and Spare Parts
N.
Read/
Write
000
R/W
001
R/W
008
R/W
100
R/W
101
R/W
102
R/W
103
R/W
104
R/W
105
R/W
106
R/W
107
R/W
108
R/W
244/301
Data
Description
Type
L
Start/Stop (in remote/ Front mode the
window is a read only)
L
Low Speed Activation
L
Remote (default) or Serial configuration
L
Soft Start (write only in Stop condition)
N
R1 Set Point type
N
R1 Set Point valve (expressed in Hz, W or s)
N
Set Point delay: time between the pump
start and the set point check (seconds)
L
Set Point signal activation type: the signal
can be "high level active" or "low level
active"
N
Set point hysteresis (in % of value)
L
Water cooling
L
Active Stop (write only in stop)
N
Baud rate
TwisTorr 304 FS AG Rack Controller User Manual / 87-901-029-01
Admitted Values
Start = 1
Stop = 0
No = 0
Yes = 1 (default = 0)
Remote = 1
Serial = 0 (default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Power
2 = Time
3 = Normal (default = 3)
4 =Pressure (available only if
the gauge is connected)
(default = 900)
0 to 99999 (default = 0)
0 = high level active
1 = low level active
(default = 0)
0 to 100 (default = 2)
0 = NO
1 = YES
0 = NO
1 = YES
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)