17
Technical Information
Programmable Analog Output Diagrams
The maximum power provided by controller to drive a gauge is 2W, if your gauge
NOTE
require more power you need an external power supply.
The controller recognizes automatically the gauge connected by
reading the signal "identification" in the connector.
Following a table with the gauge supported:
Gauge model
FRG 700/702
IMG 500
PVG 500/502
FRG 720/730
IM 500
CDG 500 T1000
CDG 500 T0100
CDG 500 T0010
CDG 500 T0001
The controller give the error "gauge error" if the gauge connected
is not supported.
There are some gauges (for example the CDG family gauge) that
have the same identification. In this case the controller
recognizes the family of the gauge but the user should select the
right gauge by front panel.
The pressure can be read by display, by serial line (win. 224)
and by Remote I/O (Programmable Analog Out), but the controller
doesn't replicate the signal from the gauge.
The Programmable analog signal is according with the following
table.
ID
1
CT-100
2
FRG-700/702
3
IMG-500
4
PVG-500/502
6
PCG-750/752
7
FRG-720/730
8
IM-500 (Eyesys Mini-IMG)
9
CDG-500 T1000
10
CDG-500 T0100
11
CDG-500 T0010
12
CDG-500 T0001
224/301
Name showed in the display
FRGa
IMG
PVGa
FRGb
IM
CDG3
CDG2
CDG1
CDG0
Gauge Model
TwisTorr 304 FS AG Rack Controller User Manual / 87-901-029-01
Pressure range
[mbar]
1.33E-2 to 1.01E3
5E-9 to 1E3
2E-9 to 1E-2
5E-4 to 1E3
5E-5 to 1.5E3
5E-10 to 1E3
1.33E-9 to 1.33E-1
2.66E+1 to 1.33E3
2.66E+0 to 1.33E2
2.66E-1 to 1.33E1
2.66E-2 to 1.33
Voltage range on
Analog Output [V]
1.11 to 8.25
1.82 to 8.6
1.8 to 8.5
1.9 to 10
0.61 to 10.23
0.774 to 10
1 to 9
0.2 to 10
0.2 to 10
0.2 to 10
0.2 to 10