1 GHz Spectrum Analyzer HM5014-2
All data valid at 23 °C after 30 minute warm-up
Frequency Characteristics
Frequency Range :
0.15 MHz...1.050 GHz
Stability:
± 5 ppm
Aging:
± 1 ppm/year
Frequency Resolution:
1 kHz (6
Center Frequency Range:
0...1.050 GHz
LO Frequency Generation:
TCXO with DDS (Digital Frequency Synthesis)
Span Setting Range:
Zero Span and 1MHz...1000MHz
(1-2-5 Sequence)
Marker:
Frequency Resolution:
1 kHz, 6
Amplitude Resolution:
0.4 dB, 3
Resolution Bandwidths
(RBW) @ 6dB:
1 MHz, 120 kHz and 9 kHz
Video Bandwidth (VBW):
4 kHz
Sweep Time
(automatic selection):
40 ms, 320 ms,1 s*)
Amplitude Characteristics (Marker Related) 150 kHz...1 GHz
Measurement Range:
-100 dBm...+10 dBm
Scaling:
10 dB / div., 5 dB/div.
Display Range:
80 dB (10 dB/div.),
40 dB (5 dB/div.)
Amplitude Frequency Response (at 10 dB Attn., Zero Span and RBW 1 MHz,
Signal – 20 dBm):
± 3 dB
Display (CRT):
8 x 10 division
Amplitude Scale:
logarithmic
Display units:
dBm
Input Attenuator Range:
0...40 dB (10 dB-increments)
Tolerance of input attenuator: ± 2 dB relative to 10 dB position
Max. Input Level (continuous)
40 dB attenuation:
+20 dBm (0,1 W)
0 dB attenuation:
+10 dBm
Max. DC Voltage:
± 25 V
Max. Reference Level:
+10 dBm
Reference Level Accuracy rel. to 500 MHz, 10 dB Attn., Zero Span and
RBW 1 MHz:
± 1 dB
Min. Average Noise Level:
approx. -100 dBm (RBW 9 kHz)
Intermodulation Ratio
typical ›75 dBc (2 Signals: 200 MHz,
rd
(3
Order):
203 MHz, -3 dB below Reference Level)
Harmonic Distortion Ratio
nd
(2
harm.):
typical › 75dBc (200MHz, Reference Level)
Bandwidth Dependent Amplitude Error rel. to RBW
1 MHz and Zero Span:
± 1 dB
Digitization Error:
±1 digit (0.4 dB) at 10 dB/div. scaling
(Average, Zero Span)
Inputs/Outputs
Measuring Input:
Input Impedance:
VSWR: (Attn. ≥ 10 dB)
Tracking Generator Output:
Output Impedance:
Test Signal Output:
Frequency, Level:
Supply Voltage for Probes (HZ 530):
Audio Output (phone):
RS-232 Interface:
Functions
Keyboard Input:
Center Frequency, Reference Level,
Tracking Generator Level
Rotary Encoder Input:
Center Frequency, Reference Level, Marker,
Tracking Generator Level
Max. Hold Detection:
Peak Value Acquisition
Quasi-Peak Detection:*
Quasi-Peak Valuation
Average:
Mean Value Acquisition
Ref. Spectrum Memory:
2 k x 8 bit
SAVE / RECALL:
Save and Recall of 10 Instrument Settings
AM demodulation:
for audio
LOCAL:
RS-232 Remote Control OFF
Readout:
Display of various Measurement Parameters
1
digit in readout)
2
1
digit,
2
digit
1
2
N socket
50 Ω
typ. 1.5:1
N-socket
50 Ω
BNC-socket
48 MHz, -30 dBm (± 2 dB)
6 V DC
3.5mm Ø jack
9pol./Sub-D
Tracking Generator
Frequency Range:
0.15 MHz...1.050 GHz
Output Level:
-50 dBm...+1 dBm
Frequency Response (0.15 MHz...1 GHz)
+1 dBm...-10 dBm:
± 3 dB
-10,2 dBm...-50 dBm:
± 4 dB
Digitization Error:
± 1 digit (0.4 dB)
Spurious Outputs
better than 20 dBc
General information
CRT:
D14-363GY, 8 x 10 div. with internal graticule
Acceleration Voltage:
approx. 2 kV
Trace Rotation:
adjustable on front panel
Power Supply:
105/253 V, 50 / 60 Hz ± 10 %, CAT II
Power Consumption:
approx. 35 W at 230V/50 Hz
Safety Class
Safety Class I (EN61010-1)
Operating temperature:
+5°C...+40°C
Storage temperature:
-20°C...+70°C
Max. rel. humidity:
5%...80% (non condensing)
Dimensions (W x H x D):
285 x 125 x 380 mm
Weight:
approx. 6.5 kg
*) in combination with software AS100E only
Accessories supplied: Line Cord, Operators Manual, HZ21 Adapter Plug
(N-plug with BNC socket) and Software for Windows on CD-ROM
Optional accessories:
HZ70 Opto-Interface (with optical fiber cable)
HZ520 Antenna
HZ530 Near Field Probe Set for EMI Diagnosis
S p e c i f i c a t i o n s