Use According To Specification; Residual Risks; Technical Specifications - Hettich ROTANTA 460 Mode D'emploi

Masquer les pouces Voir aussi pour ROTANTA 460:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN
1

Use according to specification

This machine is a medical device (laboratory centrifuge) in the sense of IVD guideline 98/79/EC and the guideline for
medical products 93/42/EEC. The centrifuge is used to separate substances or substance mixtures with a density of
max. 1.2 kg/dm³. This also includes substances and substance mixtures of human origin. The centrifuge is only
intended to be used for this purpose. A different use or application over and above this is deemed not in accordance
with the specifications. The company Andreas Hettich GmbH & Co. KG undertakes no liability for damages resulting
therefrom.
Belonging to the application according to specification is also the observance of all references contained in the
Instruction Manual and compliance with the inspection and maintenance works.
2

Residual risks

The machine is constructed according to the state of the art and the recognized technical safety regulations.
Improper use and handling can result in dangers to life and limb of the user or third parties and impairments to the
machine or to other material assets. The machine is only to be used for the specified applications and only in an
impeccable technical safety condition.
Disturbances that can interfere with the safety are to be immediately rectified.
3

Technical specifications

Manufacturer
Model
Type
Mains voltage (± 10%)
Mains frequency
Connected load
Current consumption
Cooling medium
Max. capacity
Allowed density
Speed (RPM)
Force (RCF)
Kinetic energy
Obligatory inspection (BGR 500)
Ambient conditions (EN / IEC 61010-1)
− Set-up site
− Altitude
− Ambient temperature
− Humidity
− Excess-voltage category
(IEC 60364-4-443)
− Pollution degree
Device protection class
EMC
− Emitted interference, Interference
immunity
Noise level (dependent on rotor)
Dimensions
− Width
− Depth
− Height
Weight
34/147
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
ROTANTA 460
5650
5650-01
200–240 V 1∼ 100–127 V 1∼
50 – 60 Hz
50 – 60 Hz
1000 VA
1100 VA
5.0 A
11.0 A
----
41000 Nm
Up to 2000 m above sea level
2°C to 35°C
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Not suitable for use in explosion-endangered areas.
FCC Class B
EN / IEC
61326-1,
Class B
554 mm
706 mm
715 mm
456 mm
approx. 93 kg
approx. 103 kg
ROTANTA 460 R
5660-20,
5660,
5660-70
5660-50
200–240 V 1∼
200–240 V 1∼
50 Hz
1800 VA
8.5 A
R 404A
4 x 750 ml
3
1.2 kg/dm
15000
24400
51000 Nm
yes
Indoors only
5°C to 35°C
ΙΙ
2
Ι
EN / IEC 61326-1, Class
≤ 68 dB(A)
770 mm
723 mm
706 mm
481 mm
456 mm
approx. 133 kg
5660-07
60 Hz
1900 VA
9.2 A
B

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Rotanta 460 rRotanta 460 rcRotanta 460 rf

Table des Matières