Methods
GB
CAH
0.0.0
Zero
!
Test
CAH
RESULT
RESULT
48
Hardness, Calcium with Tablet
0 – 500 mg/l
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add one CALCIO H No. 1 tablet straight from the foil to the water
sample and crush the tablet using a clean stirring rod. Dissolve the
tablet completely.
Add one CALCIO H No. 2 tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved completely.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 2 minutes.
(Countdown can be activated, see page 35)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l CaCO
Tolerances:
0 – 250 mg/l: ± 5% (full scale)
251 – 500 mg/l: ± 10% (full scale)
marks
.