Figure 7-3
Always verify and be prepared to adjust the code
on the PDM to match the code number on the test
strip vial (Figure 7-4).
It is important to always enter a code whether
using FreeStyle or FreeStyle Lite with the
OmniPod, even if some FreeStyle Lite products
indicate that no coding is required. No coding only
apply when these strips are used with certain
Abbott meters, and does not apply to OmniPod.
(Figure 7-4)
Figure 7-4
Checking Your Blood Glucose
If you need to adjust the code number after the PDM
has moved to the next screen, just press the Up/Down
Controller buttons. The code screen reappears and you
can adjust the number.
The code number remains on the PDM screen for your
reference until you have completed the control test.
From some PDM screens, you cannot access the
Freestyle® blood glucose meter. For example, you
cannot use the meter while you are activating a Pod
or when an alert, alarm or communication error
screen is displayed. In these cases, if you insert a test
strip, the PDM beeps to alert you.
3. To help you see the test strip port in reduced lighting, when
you insert a test strip into the PDM, you can turn on the port
light. Press the middle Soft Key labeled Light. To turn the light
off, press Light again.
The code number on the screen should match
the code number on the side of your test strip vial
(Figure 7-4). They must always match or your
results will be inaccurate. (See "Set the blood
glucose test strip code" later in this chapter.)
4. Wait for the PDM to display "Apply a blood sample to the strip"
on the screen (Figure 7-5).