10
Alerts and Alarms
Alarm
Occlusion detected.
Insulin delivery stopped.
Change Pod now.
Pod Error.
Insulin delivery stopped.
Change pod now.
PDM Error.
Remove Pod now.
For customer support,
refer to your User Guide.
130
Type
Source
Beep Pattern
Hazard
Pod
Continuous tone.
Alarm
and
PDM
Hazard
Pod
Continuous tone.
Alarm
Hazard
PDM
Continuous tone.
Alarm
Action
Press OK to acknowledge.
Change Pod.
Check blood glucose.
This alarm occurs when the Pod detects
an unexpected error. Press OK to
acknowledge.
Change Pod.
Check blood glucose.
This alarm occurs if an unexpected error
is detected in the PDM. Press OK to
acknowledge.
Remove Pod.
Call Customer Care immediately.
Check blood glucose.