Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
a
AM and FM radio broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment.
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the System
You can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the System, as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output power
of transmitter, in watts
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distances in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
158
Electromagnetic Immunity (Continued)
Separation distance according to frequency of transmitter, in meters
150 kHz to 80 MHz, d = 1.17 √ P
0.12
0.37
1.17
3.70
11.7
80 MHz to 800 MHz, d = 0.35 √ P
0.035
0.11
0.35
1.11
3.5
800 MHz to 2.5 GHz, d = 0.7 √ P
0.070
0.22
0.7
2.21
7.0