SYMPTOM
Alarm:
YELLOW or RED light
Illuminated.
Alarm: Continuous On RED
light
* Only applies to
SensO
models.
2
GREEN light with YELLOW
Light Flashing.
Alarm:
Rapid
Beep...Beep...
Beep...Beep
1195660-A
PROBABLE CAUSE
System Failure:
1.
Low oxygen purity.*
2.
Kinked or blocked tubing, cannula
or humidifier.*
3.
Flowmeter set at 0.5 L/min*.
4.
Unit overheating due to blocked air
intake.
5.
Insufficient power at outlet.
6.
Internal repairs required.
1.
Internal repairs required.
Potential Obstruction Alert
1.
Possible internal obstruction in the
oxygen path. Kinked or blocked
tubing, cannula or humidifier.
2.
Flowmeter set at 0.5 L/min.
SOLUTION
1.
Clean or Replace filters.
2.
Inspect for kinks or blockages. Correct, clean or replace item.
Once corrected, turn power Off for 60 seconds and then turn
power back On.
3.
Check flowmeter is set to 1.0 L/min or more. Refer to Typical
Product Parameters.
4.
a.
Remove and clean cabinet filters.
b.
Move concentrator at least 12 in (30,5 cm) from walls,
draperies, and furniture.
5.
DO NOT use extension cords. Move to another electrical
outlet or circuit.
6.
Call Service Provider.
1.
Call Service Provider.
1.
Inspect for kinks or blockages. Correct, clean or replace item.
Once corrected, turn power Off for 60 seconds and then turn
power back ON.
2.
Flowrates less than 1 L/min are not recommend.
The use of the pediatric flowmeter will deactivate the
Potential Obstruction Alert.
Troubleshooting
29