Invacare Platinum Série Manuel D'utilisation page 19

Système compatible concentrateurs d'oxygène avec senso2
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 69
CAUTION!
Risk of Damage
– Confirm the humidifier cap is not cross-threaded
on the humidifier bottle.
3.
Insert a flathead screwdriver in the plate groove B on the top
edge of the filter access door C and gently pry the filter access
door off.
Pull up and remove the humidifier bottle adapter D (next to
4.
inlet filter).
1195660-A
WARNING!
Risk of Injury
To avoid impacting the flow of oxygen:
– Install the humidifier bottle and/or other
accessories properly.
5.
Attach the humidifier bottle adapter to the humidifier bottle by
turning the wing nut on the humidifier bottle counterclockwise
until it is securely fastened.
6.
When placing the humidifier bottle in the compartment E on
the concentrator, attach the humidifier bottle/adapter tube F to
the oxygen outlet connector G on the concentrator.
When using tubing connections longer than 2.1 m (7
feet) in length, position the humidifier as close to the
patient as possible to allow for maximum humidification
output.
7.
Attach the patients' nasal cannula supply tube to the humidifier
bottle outlet H.
8.
After assembly, ensure that oxygen is flowing through the nasal
cannula.
9.
Replace the filter access door on the side of concentrator by
snapping it back into the plate groove.
Usage
19

Publicité

Table des Matières
loading

Table des Matières