Agilent Technologies Turbo-V 3K-G 969-8888 Notice De Mode D'emploi page 358

Table des Matières

Publicité

17
Technical Information
Profibus Message Mapping
The ProfiBus external diagnosis services are used to report following
unexpected situations:
Diagnostic Buffer
Tab. 15
Off set
0
1
2
3
4
5
6
358/374
Controller fail: if the controller goes in fail mode, an external
diagnostic service is required and one byte in the diagnostic
buffer signals the type of failure.
User parameter mismatch: if the user sets one or more
parameters out of their allowed range, a ProfiBus external
diagnostic service is required and some bits in the diagnosis
frame signals which parameter is wrong (i.e. over range or under
range).
Diagnostic buffer
Meaning
B
station non exist (set by master)
0
B
station not ready (slave not ready for data exchange)
1
B
cfg fault (configuration data doesn't match)
2
B
ext diag (slave has external diagnostic data)
3
B
not supported (slave doesn't support requested function)
4
B
invalid slave response ( slave sets permanent '0')
5
B
prm fault (wrong parameter assignment)
6
B
master_lock (slave is parameterised by another master)
7
B
prm req (slave has to be re-parameterised)
0
B
stat diag (static diagnosis)
1
B
fixed '1'
2
B
wd_on (response monitoring active)
3
B
Freeze mode (received freeze command)
4
B
Synch mode (received synch command)
5
B
reserved
6
B
diag deactivated (slave is parameterised by another master)
7
Reserved
Master Add (master address after parameterisation. FF= without
parameter)
Ident number high
Ident number low
External diagnosis "Header" length indication including header
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
(C)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Turbo-v 3k-g 969-8889

Table des Matières