Agilent Technologies Turbo-V 3K-G 969-8888 Notice De Mode D'emploi page 308

Table des Matières

Publicité

17
Technical Information
Technical Specification
Power supply:
Input voltage:
Input freq.:
Max input power:
Stand-by power:
Max operating power:
Protection fuse
Compliance with:
Installation Category
Pollution Degree
Power cable
Serial communication kit
Storage temperature
Environment protection
Weight (ISO 250) kg (lbs)
When the Turbo-V 3K-G has been stored at a temperature less than 5 °C, wait
NOTE
until the system has reached the above mentioned temperature before starting
the pump.
308/374
208 - 240 Vac
50 - 60 Hz
950 VA
30 to 35 W
515 W for Nitrogen and lighter gases
430 W for Argon with purge on
390 W for Argon with purge off
2 x 12,5 A
EN 55011 (Radio disturbance characteristics)
EN 61000-3-2 (Limits for harmonic current emissions)
EN 61000-3-3 (Limitation of voltage changes, voltage
fluctuations and flicker)
EN 61000-6-2 and EN 61326 (Immunity for industrial
environments) which recall:
EN 61000-4-2 (Electrostatic discharge phenomena)
EN 61000-4-3 (Radiated high-frequency phenomena)
EN 61000-4-4, EN 61000-4-5, EN61000-4-6 (Conducted high-
frequency phenomena)
EN 61000-4-11 (Voltage fluctuation, voltage dips and short
interruptions)
II
2
With European or NEMA plug 3 meters long (optional)
RS232 cable with a 9-pin D type male connector and a 9-pin
D type female connector, and T-Plus software (optional)
-20° C to +70° C
IP54
52 (114.4)
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
(C)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Turbo-v 3k-g 969-8889

Table des Matières