3 Function
The echelon comprises a hydraulic body assembly containing adjustable hydraulic valves. The
valves can be independently adjusted to increase and reduce hydraulic resistance of plantar and
dorsi flexion. The hydraulic body assembly is connected to a carrier assembly via two pivot pins.
Heel and toe springs are attached to the carrier assy using titanium screws. The foot is wrapped
in a UHM PE sock which is in turn surrounded by a PU foot shell.
4 Maintenance
Maintenance must be carried out by competent personnel.
It is recommended that the following maintenance is carried out annually:
• Remove the foot shell and glide sock, check for damage or wear and replace if necessary.
• Check all screws for tightness, clean and reassemble as necessary.
• Visually check the heel and toe springs for signs of delamination or wear and replace if
necessary. Some surface damage may occur after a period of use, this does not affect the
function or strength of the foot.
The wearer should be advised:
Any changes in performance of this device must be reported to the practitioner.
Changes in performance may include:
• Increase in ankle stiffness
• Reduced ankle support (free movement)
• Any unusual noise
Cleaning:
Use a damp cloth and mild soap to clean outside surfaces, do not use aggressive cleansers.

5 Limitations on use:

Intended life:
A local risk assessment should be carried out based upon activity and usage.
Lifting loads:
User weight and activity is governed by the stated limits.
Load carrying by the user should be based on a local risk assessment.
Environment:
The product is water proof to a maximum depth of 1 metre.
Thoroughly rinse with fresh water after use in abrasive environments such as those that may
contain sand or grit, for example, to prevent wear or damage to moving parts.
Thoroughly rinse with fresh water after use in salt or chlorinated water.
Foot products must be adequately finished to prevent water ingress into the foot shell where
possible. If water enters the foot shell, the limb should be inverted and dried before further use.
Exclusively for use between -15˚C and 50˚C
It is recommended that only Endolite products be used in conjunction with the echelon.
(5˚F to 122˚F)
.
4
938280/13-0515