IC 0T illumination controller on non-HS table
63 Recess in table top for IC 01T or IC 01T-1 illumination
control
64 sufficient space must be available underneath the table
top.
65 The illumination control is fastened to the table top with
self-adhesive tape.
3.

Non-HS illumination controllers

• An illumination controller from a foreign supplier can be
plugged into the connection (66).
• Prior calibration to the appropriate potentiometer value is
necessary.
• Controllers can be calibrated between 5 and 50 kΩ.

3.. Calibrating non-HS illumination controllers

• Remove protective cap to access connector (59).
• Connect external illumination controller (power jack
2.1 mm) and set to maximum illumination (turn adjustment
knob to right until limit stop).
• Switch on PS-LED power supply unit.
• Press button S1 for longer than 1 second. The device will
change over to calibration mode. The orange LED (L1)
indicates that the illumination controller is being calibrated.
• After approx. 1 to 3 seconds (depending on the type of illu-
mination controller) the calibration process will be com-
plete. This will be indicated by two short blinks of the green
LED (L1).