Safety
Ambient conditions
• Transportation Temperature
Air pressure
Relative humidity
• Storage
Temperature -
Air pressure
Relative humidity
• Working
Temperature
Air pressure
Relative humidity
Installation
• To avoid condensation, allow the instrument to adjust to
room temperature for several hours before unpacking.
• Always mount the slit lamp and the headrest on an electric
insulated and fire resistant table top.
Operation and surrounding
• Only qualified and trained personnel should operate the
equipment, the training is at the owner's responsibility.
• Use only Haag-Streit accessories.
• Shut down after every use.
• Never operate the instrument in the same room with com-
bustible gases, volatile solvents (alcohol, benzol) or flam-
mable anesthetic agents.
• Avoid humidity.
Electrical
• Only a hospital grade 3-conductor electrical power supply
cable must be used. For USA and Canada: Detachable
Power Supply Cord Set, UL Listed, type SJE, SJT or SJO,
3-conductor, not smaller than 18 AWG.
• Plug, cable and ground lead connection of the socket have
to be in perfect condition.
To avoid the risk of an electric shock, the-
se devices should only be connected to a
supply network with an earth wire.
Cleaning
• Clean the housing with a water dampened cloth only.
• No liquids or corrosive agents.
Warranty / product liability
• This instrument should be handled in accordance with the
'Safety' chapter. Improper handling may damage the in-
strument. Any resultant claim made under the guarantee
will not be accepted.
• Continued use of an instrument that has been damaged
by incorrect handling may lead to personal injury. The ma-
nufacturer cannot accept liability in this case.
• Repairs and alterations to this equipment should only be
carried out by Haag-Streit service technicians or by autho-
rized persons.
-40 °C to +70 °C
500 hPa to 1060 hPa
10% to 95%
10 °C to +55 °C
700 hPa to 1060 hPa
10% to 95%
+10 °C to +35 °C
800 hPa to 1060 hPa
30% to 75%
1500 7220243.04000
© Haag-Streit AG, 3098 Koeniz, Switzerland