Clean and disinfect the full-face mask as follows:
1. Remove any accessories and clean them separately, if required
2. Clean all components with lukewarm water and Sekusept
Cleaner with a soft cloth (max. temperature: 30 °C; concentration
depending on the degree of contamination: 0.5 - 1 %)
3. Rinse all parts thoroughly under running water.
4. Prepare a disinfectant bath made of water and Incidin
(max. temperature: 30 °C; concentration: 1.5 %)
5. Put all components that need disinfection into the disinfection bath
for 15 minutes.
6. Rinse all parts thoroughly under running water.
7. Allow all parts to air-dry or be dried in the drying cabinet
(temperature: max. 60 °C). Protect against direct sunlight.
8. Mount the accessories, if necessary.
NOTICE
i
i
Full-face masks with PC visor can also be cleaned and
disinfected in the washing machine. Information can be
obtained from Dräger.
4.3
Tests and inspections
Inspect the mask in accordance with the inspection and maintenance
intervals.
4.3.1
Visual inspection of the speech diaphragm
The speech diaphragm and the O-ring must be clean and
undamaged; if not, clean or replace (see chapter 4.4.4 on
page 20).
1)
Follow the relevant instructions for use.
®
2)
Sekusept
is a registered trademark of Ecolab Deutschland GmbH
®
3)
Incidin
is a registered trademark of Ecolab USA Inc.
18
4.3.2
1)
.
The inhalation valve is visible through the visor in the equipment
®
connector. Blow into the equipment connector to check whether
both wings of the valve disc move freely. Replace the valve disc, if
2)
.
necessary (see chapter 4.4.6 on page 21).
®
Rapid
4.3.3
3)
.
1. Pull off the exhalation valve cover from
the equipment connector.
2. Press one arm inwards and remove the
spring holder.
3. Remove the valve disc from the guide.
4. The valve disc and the valve seat must
be clean and undamaged. Clean or
replace if necessary.
5. Moisten the valve disc with water and refit. The valve disc must
slide into the guide under its dead weight.
6. Fit the spring holder and make sure that both arms engage at the
sides. It is marked with "L" = left and "R" = right and fits at an angle.
(approved by Dekra Exam)