Technical Information

Vibration Isolator Installation

Vibration Isolator Installation

Five vibration isolators for ISO and CFF inlet flange version pumps
are available as accesso-ries.
The five model part numbers are the following:
They typically reduce the vibration transmitted from the Turbo-V
750/850 TwisTorr Pumping System to the system by a factor of 20.
Please refer to the relevant instruction manual for installation.

Connection of the High Vacuum Flange

In the case the rotor is suddenly blocked, the torque arising from the
system and the high vacuum flange must be absorbed. Only the
components listed in the following can be used to fasten the pump to
the high vacuum flange.
WARNING!
It is mandatory to connect the pump in such a way to withstand the torque
specified in the "IN-STRUCTION FOR USE" section. Specifically pay attention
to clamp design, material of clamps and bolts and bolt fixing torque.
382/406
model 969-9345 for ISO 160 flange;
model 969-9346 for ISO 200 flange;
model 969-9334 for CFF 6" flange;
model 969-9335 for CFF 8" flange;
model 969-9336 for CFF 10" flange.