Setting:
Admissible alteration of full scale
output, variable sensitivity, adjust-
able with potentiometer P200.
Setting range:
0,85 - 1,15 · I
100%
0
– Remove calibration seal (small sticker).
– Connect the power supply
– Connect an average burden and a meter (e.g.
MetraHit 18s) to output
– Calibrate the output "zero point" using P201. (The "zero
point" corresponds to the demand value for zero input).
– P201 does not influence the full-scale value.
– Connect the DC calibration voltage of 5 V to KP1 (–) and
KP2 (+).
– Adjust the calibration voltage so that the output signal
indicates 100% ± 0.1%.
– The calibration voltage thus determined equals the
reference voltage U
.
ref actual
– The available setting range is
U
= U
· 0,85 - 1,15.
ref demand
ref actual
– Recalibrate the output signal to 100% ± 0.1% using P200.
– Replace the seal over the calibration openings.
Examples of the max. setting:
Alteration of measuring range from 0 - 1.2/6 A to
0 - 1.38/6.9 A
Measuring range: 0 - 1.2/6 A,
Reference voltage determined U
Measuring range: 0 ... 1.38/6.9 A, Output: 0 -- 20 mA
1.38
U
= 4.982 ·
= 5.729 V
ref demand
1.2
U
= 5.729 V is the new calibration voltage
ref demand
U
= Old calibration voltage
ref actual
U
= New calibration voltage
ref demand
I
= Full-scale of measurement range on delivery
N
P201
10 11 12 13
+
4-20 mA
Fig. 7. Locations of the potentiometers P201 and P200 and the
terminals KP1 and KP2.
(± 15%)
N
I
N
0
Input
0.85
1
1.15
.
.
Output: 0 - 20 mA
= 4.982 V
ref actual
P200
KP1 (–) KP2 (+)
0,5

7. Commissioning and maintenance

Switch on the power supply and the measuring input. It is
possible during the operation to disconnect the output line
and to connect a check instrument, e.g. for a functional test.
No maintenance is required.

8. Releasing the transducer

Release the transducer from a top-hat rail as shown in Fig. 8.
Fig. 8

9. Instruments admissions

Germanischer Lloyd
Type approval certificate: 12 258-98 HH
CSA approved for USA and Canada
file-nr. 204767
FCC Compliance and Canadian DOC Statement
This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to both part
15 of the FCC Rules and the radio interference regulations
of the Canadian Department of Communications: These
limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses
and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may
cause harmful interference to radio communications.
Operation of this equipment in a residential area is like to
cause harmful interference in which case the user will be
required to correct the interference at his own expense.
11

Publicité

Chapitres

Table des Matières
loading

Table des Matières