Dräger Aerotest HP Notice D'utilisation page 29

Masquer les pouces Voir aussi pour Aerotest HP:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
Table of measurement recommendations/limit values for MultiTest
Measurements in air
Pressure: 3 bars ±0.2
Measuring point
in the
Tube
measuring
device for
Oil
Oil 10/a-P
Impactor
CO
CO
100/a-P 500 ppm
2
2
SO
SO
0.5/a
2
2
NO
NO
0.2/a
x
x
CO
CO 5/a-P
H
O
H
O 20/a-P
2
2
Measurements in O
2
Pressure: 3 bars ±0.2
Measuring point
in the
Tube
measuring device
for
CO
CO
100/a-
2
2
P
CO
CO 5/a-P
H
O
H
O 20/a-P
2
2
Flow
Test time
Limit value
(l/min)
±10 %
3
0.1 mg/m
4
depending on oil type
see IfU for Oil 10/a-P
3
0.1 mg/m
4
5 min
0.2
1 ppm
0.2
1 ppm
0.2
2 ppm
0.2
5 ppm
0.2
60 ppm
4
3
(45 mg/m
)
Flow
Test time
Limit value
(l/min)
±10 %
300 ppm
0.2
5 ppm
0.2
60 ppm
4
3
(45 mg/m
)
(min)
Display
Measurement range
see instructions for use for oil tube
see instructions for use for impactor
5
ppm
100 to 3000 ppm
5
Evaluation on scale of n = 10
ppm SO
1 to 25 ppm
2
10
Evaluation on scale of n = 20
Display x
0.25 to 1 ppm
0.5 = ppm
(only valid for scale
SO
range 0.5 to 2 ppm)
2
2.5
ppm NO
0.2 to 6 ppm
x
5
ppm
5 to 150 ppm
3
10
mg/m
20 to 100 mg/m
(min)
Display
Measurement range
5
ppm
100 to 3000 ppm
5
ppm
5 to 150 ppm
3
10
mg/m
20 to 100 mg/m
Evaluation
Deviation from
standard
±10 to 15 %
15 % ±20 %
±25 %
±30 %
±10 to 15 %
3
±10 to 15 %
Evaluation
Deviation from
standard
±10 to 15 %
±10 to 15 %
3
±10 to 15 %
29