Maximum Timer Value (Timer Max); Set Working Parameters; Target Speed (Target Rpm); Timer Function - IKA ULTRA_TURRAX Mode D'emploi

Table des Matières

Publicité

Timer maximum value (timer max)
11.4
The default "timer max" is 30 minutes in the device. User can
change the setting in 10 second intervals and up to 30 minutes
in the main menu.
The "timer max" setting will be stored automatically when opera-
tion is over or the device is switched off.
• Set "timer max"
 Press the "Menu/Back" button (C) in the main menu.
 Turn the "Rotating/Pressing" knob (B) to select the "timer max"
in the main menu.
 Press the "Rotating/Pressing" knob (B)
 A mark
will appear to indicate the activated item.
 Turn the "Rotating/Pressing" knob (B) to change the value.
 Press the "Rotating/Pressing" knob (B) again to complete the
process
 Press the "Menu/Back" button (C) to return to the working
screen.
Note: The timer „timer max" can be adjusted with the "Rotating/
Pressing" knob (B) in a range from 10 seconds to 30 minutes. This
is achieved in magnitudes of 10 seconds. In remote control mode
via computer, adjusting magnitude is 1 second.
12
Setting working parameters
12.1

Target speed (target rpm)

Set "target rpm" within "rpm max" range in the working screen
by rotating the "Rotating/Pressing" knob (B). Adjustment magni-
tude is 10 rpm. In remote control mode via computer, adjusting
magnitude is 1 rpm.
• Setting "target rpm" in working screen
 Turn the "Rotating/Pressing" knob (B) to change the value.
 Press "Rotating/Pressing" knob (B) to start working.
 Rotation symbol ( ,,) indicates running.
12.2

Timer function

Set "Timer" within "timer max" range in the working screen by ro-
tating the "Rotating/Pressing" knob (B). Adjustment magnitude is 10
seconds. In remote control mode via computer, adjusting magnitude
is 1 second.
If the "Timer" is deactivated when the device is started, the device
will automatically count from 0 to 30 minutes and then the unit will
stop running.
• Setting "Timer" value (timer sec) in working screen
 Press "Timer" button (D) to activate the timer setting in the
working screen.
 Turn the "Rotating/Pressing" knob (B) to change the value.
 Press "Timer" button (D) again to complete the setting.
 Current values are stored.
24

Publicité

Table des Matières
loading

Table des Matières