Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY –
MIO CARE PRO is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Impluses
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supplì
input lines
IEC 61000-4-11
IACER Srl
FOR ALL EQUIPMENT AND SYSTEMS
Test level
Compliance level
IEC 60601
±8kV in contact
±8kV in contact
±2Kv, ±4kV,
±2Kv, ±4kV,
±8kV; +15kV on
±8kV; +15kV on
air
air
±2kV
±2kV
for power
for power
supplies lines
supplies lines
±1kV
±1kV
for input/output
for input/output
lines
lines
±0.5kV, ±1kV
±0.5kV, ±1kV
Line(s) to line(s)
Line(s) to line(s)
±0.5kV, ±1kV,
±0.5kV, ±1kV,
±2kV
±2kV
Line(s) to earth
Line(s) to earth
0% U
a 0°, 45°,
% U
T
90°, 135°, 180°,
90°, 135°, 180°,
225°, 270° and
225°, 270° and
315°
for
0,5
315°
cycles
cycles
0% U
0% U
T
for 1 cycle and
for 1 cycle and
70% U
70% U
T
for 25/30 cycles
for 25/30 cycles
at singular phase
at singular phase
0% U
for
0% U
T
250/300 cycles
250/300 cycles
56
Electromagnetic
environment - guide
Floors
wood, concrete or
ceramic tile. If floors
are
synthetic
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
a 0°, 45°,
Main power quality
T
should be that of a
typical commercial or
for
0,5
hospital environment.
If the user of MIO
CARE PRO requires
continued operation
T
during power mains
interruptions,
T
recommended
MIO CARE PRO be
powered
uninterruptible
for
power supply or a
T
battery.