Did you bend or move the test strip while the test was in
progress?
Does the concentration table belong to the test strips that are
currently in the meter?
Did you perform the test within the correct temperature range
(+10 to +40 °C)?
Has the use-by date for the control solution or the test strips
expired (see label on the bottle and the test strip drum next to
the
symbol)?
Has the 90-day use-by period for the test strip drum expired
(see Chapter 2.3.1 and possibly the clock symbol
meter's display)?
Are the test strip guide and measuring window clean?
Have you observed the proper storage conditions for the
meter, test strips and control solution (see Chapter 11 and the
package inserts that came with the test strips and control
solution)?
Has the control solution been open for less than three
months? Once opened, control solutions are only stable for
three months and must not be used after this period.
138
no
yes
yes
no
no
in the
yes
yes
yes
Checking your meter